Customization: | Available |
---|---|
After-sales Service: | Online Technical Support |
Warranty: | 2 Years |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Infitek REV-2000 Series 2L Laboratory OEM Laboratory Vacuum Rotary Evaporator
The motor of REV-2000series has 0-180RPM step-less speed regulation and automatic lifting function, the main body uses the aluminum alloy material, smooth, safe, and beautiful. Revolving body is consisted of a set of stable and reliable micro motors, speed motors and integrated circuit control devices. In the role of electronic speed, it can be rotated stably in 0-180RPM range.
Characteristics
REV-2000A: temperature digital display
REV-2000B: temperature and speed dual-digital display
REV-2000C: temperature and speed LCD display
Feature:
Model | REV-2000A | REV-2000B | REV-2000C |
Glass material | GG-17 | ||
Stent material | Aluminium alloy | ||
Bath material | Teflon composite pot, fully enclosed heating | ||
Rotary flask volume | 0.5-2L 2# standard mouth | ||
Collecting flask volume | 1L 35# ball mill mouth | ||
Rotary power | 40W | ||
Rotary speed | 0-180rpm | ||
Heating power | 1.5 kW | ||
Temperature control of bath | Room Temperature~99ºC | ||
Temperature precision | ±1ºC | ||
Elevating power | 15W | ||
Elevating stroke | 120 mm | ||
Voltage/frequency | 220V/50Hz | ||
Exterior size (L×W×H) mm | 660×420×830 | ||
Packing size (L×W×H) mm | 630×560×540 | ||
Packing weight (kg) | 26 |
1.What is the mechanism of manual lifting and auxiliary lifting of manual rotary evaporator?
The manual lifting part is a mechanical constant force lifting mode, which is safe, reliable, and low failure; the auxiliary lifting part is a stepless mechanical positioning adjustment.
2.Is it possible for a small rotary evaporator with a 5L evaporating flask?
It is not recommended to use this way, there is a risk. The excessive load of the glass shaft will cause the bearing parts to wear and reduce the life, which will affect the sealing performance. In severe cases, the shaft tube may be broken.
3.REV100-Pro can only rotate at a maximum of 280rpm when the forward and reverse rotation is set, and only 200rpm for one-way rotation. Why?
One-way rotation can reach 280rpm, no problem. To ensure safety, the maximum speed is limited to 200rpm.
4.Is it better to put the rotary steaming system in a fumehood?
It is better to put the rotary steaming system in a fume hood, and the taste can be discharged directly outside.
5.Is magnetic inspection required for air transportation?
There is a lifting motor, which requires magnetic inspection.
6.How to connect the rotary steaming and vacuum controller, and what is the function of the vacuum pump?
The vacuum controller is connected between the rotary steamer and the vacuum pump, in series. When using a rotary evaporation evaporator, the vacuum degree is not appropriate, and the reagents will suddenly boil and bump. By using a vacuum controller, in series with the rotary evaporation, the internal vacuum of the rotary evaporation can be accurately controlled, especially for the concentration of volatile reagents, which can achieve the maximum The most ideal recovery rate. It is also widely used in the vacuum distillation of organic reagents, the gradual control of the vacuum degree, and the gradual extraction of solvents.