Model | PH-B500T | |
Parameters | pH/Temp.(mV/ORP) | |
pH | Range | -2.000 to 20.000 pH |
Resolution | 0.1, 0.01, 0.001pH | |
Accuracy | ±0.002 pH | |
Calibration Points | Up to 6 | |
Standard Customization | Yes | |
Calibration Reminder | Yes | |
Standard Recognition | NIST, GB, DIN, USA, MERK, JIS buffers | |
Slope Limit | Yes | |
mV/ORP | Range | -2000.00 to 2000.00 mV |
Resolution | 0.1, 0.01 mV | |
Accuracy | ±0.1 mV or ±0.03% | |
EH ORP Mode | Yes | |
ORP Calibration Points | 1 custom point (relative mV) | |
Temperature | Range | -10 to 135 ºC, 14 to 275 ºF |
Unit | ºC, ºF | |
Resolution | 0.1 | |
Accuracy | ±0.1 | |
Measurement | Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous |
Reading Prompts | Reading, Stable, Locked | |
Temp. Compensation | ATC, MTC | |
Data management | Data Storage | 1000 Groups |
GLP Features | Yes | |
Log Management | Yes | |
Inputs | pH Electrode | BNC(Q9) |
Temp. Probe | 4-pin aviation connector | |
Outputs | USB | USB 2.0 flash memory device, PC |
RS 232 | printer, scanner, auto-sampler | |
Display options | Backlight | Yes |
Auto Shutdown | 1~60 min, off | |
IP Rating | IP54 | |
Date and Time | Yes | |
General | Electricity | AC Adapter, 100-240V AC input, DC24V output |
Dimensions | 220×195×68 mm | |
Weight | 950g |
Q: How about quality of the goods?
A:
Assured availability of parts supply for 10 years at least
Q: Is OEM service available.
A: Most of the running products are available for OEM customization, including logo, package and function, if attainable, customization.
Q: How long is the lead time:
A: Generally, the lead time for most stocked products is about 7~15 days, vary by models and production schedule. For those non-stocked products, the manufacturing lead time is around 20 days.