Overview

Basic Info.

Model NO.
SP-LUV7600
Wavelengths
UV Spectrometer
Certification
ISO9001, CE, FDA
Transport Package
High-Quality Cartons/Crates
Trademark
Infitek
Origin
China
Production Capacity
200 Units/Month

Product Description

Infitek Double Beam UV-Visible Spectrophotometer with Ultra-Low Stray Light with CE FDA
Infitek Double Beam UV-Visible Spectrophotometer with Ultra-low stray light
Features

1. Continuously variable spectral bandwidth:
The instrument's spectral bandwidth is continuously variable from 0.5nm to 6nm, the minimum bandwidth is 0.5nm, and the variable interval is 0.1nm, which not only ensures excellent spectral resolution, but also provides a variety of bandwidth options, which can better match the analysis and test targets.
2. Ultra-low stray light:
Excellent C-T monochromator optical system, advanced electronic system, to ensure ultra-low stray light level better than 0.03%, to meet the user's measurement needs of high absorbance samples.
3. High-quality devices:
The core devices are made of high-quality imported parts to ensure the stability and longevity of the instrument. For example, the core light source device is derived from the long-life lamp of Hamamatsu in Japan, which guarantees a working life of more than 2000 hours, greatly reducing the maintenance frequency and cost of daily replacement of the light source of the instrument.
4. Long-term stability and reliability:
The design of the optical dual-beam optical system, coupled with real-time digital proportional feedback signal processing, effectively offsets the signal drift of light sources and other devices, ensuring the long-term stability of the instrument baseline.
5. High wavelength accuracy:
The high-level wavelength scanning mechanical system ensures the accuracy of wavelengths better than 0.3nm and the repeatability of wavelengths better than 0.1nm. The instrument uses the built-in spectral characteristic wavelengths to automatically perform wavelength detection and correction to ensure long-term wavelength accuracy stability.
6. Light source replacement is convenient:
The instrument can be replaced without removing the shell. The light source switching mirror supports the function of automatically finding the best position. The in-line  tungsten lamp design does not require optical debugging when replacing the light source.
7. Instrument is rich in functions:
The instrument is equipped with a 7-inch large-screen color touch LCD screen, which can perform wavelength scanning, time scanning, multi-wavelength analysis, quantitative analysis, etc., and supports the storage of methods and data files. View and print the map. Easy to use, flexible and efficient.
8. Powerful PC software:
The instrument is connected to the computer via USB. The online software supports multiple functions such as wavelength scanning, time scanning, kinetic testing, quantitative analysis, multi-wavelength analysis, DNA / RNA analysis, instrument calibration, and performance verification. Support user authority management, operation traceability, and meet various requirements in different analysis fields such as pharmaceutical companies.

Specifications
SP-LUV7600
Optical system Optical double beam system 
Monochromator system Czerny-Turner monochromator 
Grating 1200 lines / mm high-quality holographic grating
Wavelength range 190nm~1100nm
Spectral bandwidth 0.5~6.0nm
Wavelength accuracy ±0.3nm 
Wavelength reproducibility ≤0.1nm
Photometric accuracy ±0.002Abs(0~0.5Abs),±0.004Abs(0.5~1.0Abs),±0.3%T(0~100%T) 
Photometric reproducibility ≤0.001Abs(0~0.5Abs),≤0.002Abs(0.5~1.0Abs),≤0.1%T(0~100%T)
Stray light  ≤0.03%(220nm,NaI;360nm,NaNO2)
Noise ≤0.1%T(100%T),≤0.05%T(0%T),≤±0.0005A/h(500nm,0Abs,2nm bandwidth)
Baseline flatness  ±0.0008A 
Baseline noise ±0.1%T 
Baseline stability  ≤0.0005Abs/h
Modes  T/A/Energy
Photometric range  -0.00~200.0(%T)   -4.0~4.0(A)
Scan speed  High / medium / low / very low
WL scan interval 0.05/0.1/0.2/0.5/1/2 nm
Light source  Hamamatzu Lamp,Osram halogen tungsten lamp
Detector  Photocell
Display 7-inch large-screen color touch LCD screen
Interface  USB-A/USB-B
Power AC90V~250V,  50H/ 60Hz
Dimension, Weight 600×470×220mm, 18Kg
Packging and Shipping

 

Infitek Double Beam UV-Visible Spectrophotometer with Ultra-Low Stray Light with CE FDA
Certifications
Infitek Double Beam UV-Visible Spectrophotometer with Ultra-Low Stray Light with CE FDA

 

 

About us

Infitek Double Beam UV-Visible Spectrophotometer with Ultra-Low Stray Light with CE FDA

Infitek Double Beam UV-Visible Spectrophotometer with Ultra-Low Stray Light with CE FDA
FAQ

Q: How about quality of the goods?

A:lAll the running products comply with CE standards.

lIndependent quality inspection team, conducting multiple inspections and random inspections from parts supply to product dispatch.

lWell-known third-party testing service available, such as SGS, BV, and TUV, etc.

Q: What are your pre- and after-sales services?

A:lTimely reply to pre-sales enquiries

lReal-time update of manufacturing and shipment status

lContracted third-party maintenance company in over 60 countries

   lAssured availability of parts supply for 10 years at least

Q: Is OEM service available.

A: Most of the running products are available for OEM customization, including logo, package and function, if attainable, customization.

Q: How long is the lead time:

A: Generally, the lead time for most stocked products is about 7~15 days, vary by models and production schedule. For those non-stocked products, the manufacturing lead time is around 20 days.
Infitek Double Beam UV-Visible Spectrophotometer with Ultra-Low Stray Light with CE FDA

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now