1. 12.1″ color TFT screen
2. Multi‐channel display
3. Big font interface
4. ST and arrhythmia analysis
5. Drug dose calculation and titration table
6. OxyCRG dynamic view display
7. Respiration monitoring with apnea alarms
8. Alarm review, arrhythmia review
9. Patient info. Input
10. Anti‐defibrillation design
11. NIBP over pressure protection
12. Plug‐in rechargeable battery, AC/DC available.
13. Supporting wire or wireless network.
CG | Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL,V) |
Lead selection: I, II, III, avR, avL, avF, V | |
Wavef orm: 2 ch | |
Lead mode: 3 Leads ( R, L, F or RA, LA, LL) | |
Lead selection: I, II, III, | |
Wave form: 1 ch | |
Gain:2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV, auto | |
HR and Alarm Range:Adult 15 ~ 300 bpm | |
Neo/Ped 15 ~ 350 bpm | |
Accuracy: ±1% or±1bpm, which is greater | |
Resolution:1 bpm | |
Sensitivity:> 200 (uVp-p) | |
Differential Input Impedance:> 5 MΩ | |
CMRR | Monitor > 105 dB |
Operation > 105 dB | |
Diagnosis > 85 dB | |
Electrode offset potential: ±300mV | |
Leakage Current: < 10 uA | |
Baseline Recovery: < 3 S After Defi. | |
ECG Signal Range: ±8 m V ( Vp-p ) | |
Bandwidth: Surgery 1 ~ 20 Hz | |
Monitor 0.5 ~ 40 Hz | |
Diagnostic 0.05 ~ 130 Hz | |
Calibration Signal: 1 (mVp-p), Accuracy:5% | |
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV | |
ARR Detecting: Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,BRADY, MISSED BEATS, | |
PNP, PNC,Alarm Available,Review Available | |
Respiration | Method :Impedance between R-F(RA-LL) |
Differential Input Impedance:> 2.5 MΩ | |
Measuring Impedance Range: 0.3~3 Ω | |
Base line Impedance Range: 200~4 KΩ | |
Bandwidth: 0.1~2.5Hz | |
Resp. Rate:Measuring and Alarm Range | |
Adult 0 ~ 120 Brpm | |
Neo/Ped 0 ~ 150 Brpm | |
Resolution 1 Brpm | |
Accuracy ±2 Brpm | |
Apnea Alarm:10 ~ 40 S | |
NIBP | Method: Oscillometric |
Mode: Manual, Auto, STAT | |
Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min) | |
Measuring Period in STAT Mode 5 Min | |
Pulse Rate Range : 40 ~ 240 bpm | |
Alarm Type :SYS, DIA, MEAN | |
Measuring and alarm range: Adult Mode | |
SYS 40 ~ 270 mmHg | |
DIA 10 ~ 215 mmHg | |
MEAN 20 ~ 235 mmHg | |
Pediatric Mode | |
SYS 40 ~ 200 mmHg | |
DIA 10 ~ 150 mmHg | |
MEAN 20 ~ 165 mmHg | |
Neonatal Mode | |
SYS 40 ~ 135 mmHg | |
DIA 10 ~ 100 mmHg | |
MEAN 20 ~ 110 mmHg | |
Resolution: 1mmHg | |
Accuracy:Maximum Mean error ±5mmHg | |
Maximum Standard deviation ±8mmHg | |
Overpressure Protection: Adult Mode 297±3 mmHg | |
Pediatric Mode 240±3 mmHg | |
Neonatal Mode 147±3 mmHg | |
SpO2 | Measuring Range: 0 ~ 100 % |
Alarm Range: 0 ~ 100 % | |
Resolution: 1 % | |
Accuracy:70% ~ 100% 2 % | |
0% ~ 69% unspecified | |
Actualization interval about 1 Sec. | |
Alarm Delay: 10 Sec. | |
Pulse Rate: Measuring and Alarm Range 20~300 bpm | |
Resolution:1bpm | |
Accuracy ±3bpm | |
TEMP | Channel: 2 |
Measuring and Alarm Range 0 ~ 50 ºC | |
Resolution: 0.1 ºC | |
Accuracy : ±0.2 ºC | |
Actualization intervalabout 1 Sec. | |
Average Time Constant: < 10 Sec. | |
IBP | Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2 |
Measuring and alarm range:ART 0 ~ 300 mmHg | |
PA -6 ~ 120 mmHg | |
CVP/RAP/LAP/ICP-10 ~ 40 mmHg | |
P1 / P2-10 ~ 300 mmHg | |
Press Sensor: Sensitivity 5 uV/V/mmHg | |
Impedance 300-3000Ω | |
Resolution: 1 mmHg | |
Accuracy:±2% or 1mmHg which is greater | |
Actualization interval : about 1 secretary | |
Standard: ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature | |
Optional: Nellcor SpO2, EtCO2, IBP,TEMP-2, Thermal Recorder,Wall Mount,Trolly,Central station |
Q: Are you a trading company or a manufacturer?
A: Based on independent manufacturing capability, we also integrate both domestic and international resources providing one-stop service for you all.
Q: How about quality of the goods?
A:
l All the running products comply with CE standards.
l Independent quality inspection team, conducting multiple inspections and random inspections from parts supply to product dispatch.
l Well-known third-party testing service available, such as SGS, BV, and TUV, etc.
Q: What are your pre- and after-sales services?
A:
l Timely reply to pre-sales enquiries
l Real-time update of manufacturing and shipment status
l Contracted third-party maintenance company in over 60 countries
Assured availability of parts supply for 10 years at least
Q: Is OEM service available.
A: Most of the running products are available for OEM customization, including logo, package and function, if attainable, customization.
Q: How long is the lead time:
A: Generally, the lead time for most stocked products is about 7~15 days, vary by models and production schedule. For those non-stocked products, the manufacturing lead time is around 20 days.